ESI-MS confirms the molecule is what the synthesis claims.
Electrospray ionisation in positive mode, observed molecular weight matched against the theoretical sequence within ±1 Da. The assay HPLC can't replace — and the one truncations and adducts can't hide from.
Six steps from sample aliquot to identity confirmed.
Mass spec is a confirmatory assay — it answers a question HPLC can't. The same lot that cleared the chromatographic purity floor goes through this path before release is signed.
-
01
Aliquot & dilution
20 µg of the same sample analysed by HPLC is diluted in 50:50 acetonitrile / water + 0.1% formic acid to a working concentration of 5 µg/mL — the sweet spot for stable electrospray without source overload.
-
02
Calibration check
Instrument calibrated against a polypeptide reference (cesium iodide cluster ions or a sequenced standard), m/z accuracy verified at < 5 ppm across the full scan range. No sample is run on a drifted instrument.
-
03
Direct infusion (or LC-MS)
Sample loaded into a glass syringe and infused at 5 µL/min through the ESI source. For complex samples, an in-line HPLC column is used in MS-compatible mobile phase (formic acid replaces TFA) for inline desalting.
-
04
Full-scan acquisition
Positive electrospray, capillary 4.0 kV, source 150 °C, cone 40 V. Scan range m/z 100–3000 at 1 spectrum/sec, averaged over 60 seconds for signal stability and isotope envelope clarity.
-
05
Charge-state deconvolution
Multi-charged peaks (typically [M+4H]<sup>4+</sup> through [M+12H]<sup>12+</sup> for a 9 kDa peptide) are deconvoluted to a single neutral molecular weight by maximum-entropy algorithm. Output: one MW value with isotope-resolved precision.
-
06
Identity decision
Observed MW compared to the theoretical MW calculated from the sequence. ≤ ±1 Da deviation: identity confirmed, the lot moves to LAL endotoxin testing. Larger deviation: lot is held and the synthesis batch is investigated.
A multi-charge envelope, deconvoluted to one number.
Electrospray produces a series of multi-protonated ions of the same molecule. The pattern of the envelope is itself a confirmation: a clean, predictable distribution means a single defined species. Here is what to read in one.
Charge envelope
Eight protonation states from <b>[M+12H]<sup>12+</sup></b> down to <b>[M+4H]<sup>4+</sup></b>, distributed in a clean Gaussian envelope. The shape itself confirms a single defined molecule; a chimera or contamination would distort it.
Deconvoluted MW
Maximum-entropy algorithm reduces the multi-charge envelope to one neutral mass: <b>9111.2 Da</b>. Theoretical: 9111.5 Da. Δ = <b>−0.3 Da</b> — well inside the ±1 Da release window.
Baseline
Flat between charge states, no parasitic ions, no Na<sup>+</sup>/K<sup>+</sup> adduct ladders, no early-eluting matrix interference. The signal-to-noise of the most abundant peak exceeds 200:1.
Source, analyser, scan, calibration.
Mass spectrometry without disclosed parameters is just a number with a stamp. Here is exactly what the lab runs against — reproducible on any modern Q-TOF or Orbitrap-class instrument.
Ionisation source
| Mode | Electrospray, positive (ESI+) |
|---|---|
| Capillary voltage | 4.0 kV |
| Cone voltage | 40 V |
| Source temperature | 150 °C |
| Desolvation gas | N₂, 800 L/h, 350 °C |
| Flow rate | 5 µL/min (direct infusion) |
Mass analyser
| Type | Q-TOF, high resolution |
|---|---|
| Resolution | ≥ 25 000 FWHM |
| Mass accuracy | < 5 ppm RMS |
| Scan range | m/z 100 – 3000 |
| Scan rate | 1 spectrum/sec |
| Acquisition | 60 s, signal-averaged |
Calibration & QC
| Calibrant | Sodium formate cluster ions |
|---|---|
| Cadence | Daily lock-mass + per-batch verify |
| Drift rejection | > 5 ppm = no run |
| System suitability | Reference peptide, m/z and intensity |
| Replicates | 2 acquisitions per lot |
| Reference standard | Independent, lot-traceable |
Sample & processing
| Diluent | 50:50 ACN/H₂O + 0.1% formic acid |
|---|---|
| Concentration | 5 µg/mL working solution |
| Charge states processed | +4 to +12 typical |
| Deconvolution | MaxEnt, Bayesian model |
| Output mass tolerance | ± 0.1 Da resolved |
| Reporting | Δ vs. theoretical MW, in Da |
What an MW deviation actually means.
A peptide is its sequence; a sequence has a defined mass. A deviation between observed and theoretical MW is a signal — sometimes negligible, sometimes a synthesis failure. Here is how we read it.
Where most of our lots actually land. Within isotope envelope precision — the molecule matches the sequence atom-for-atom.
Within the ±1 Da release window. Likely calibration noise on a large peptide; sequence integrity is intact.
Outside the release window but inside common modification masses. Re-run on a freshly calibrated instrument; if confirmed, lot is held.
Sequence error, missing residue, gross modification, or wrong vial labelled. Rejected immediately and the synthesis batch is investigated.
What every mass-shift in the spectrum means.
Just like HPLC tells a story in retention shifts, ESI-MS tells one in mass shifts. The same secondary peaks recur across syntheses — and each one points at a specific failure mode.
Four conditions that reject identity.
HPLC purity above 99% does not save a lot from MS rejection. These are the conditions under which the QC officer signs an MS rejection — independent of the chromatographic result.
Deconvoluted MW deviation > ±1 Da
The release window. A 9 kDa peptide reading 9113 instead of 9111.5 is not "close enough" — it's a +1.5 Da shift consistent with deamidation, and the lot is held until the source of the shift is identified.
Secondary species > 1% of main
An impurity peak in the deconvoluted spectrum at ≥ 1% of the main MW signal indicates a modification or truncation that HPLC could not separate. The lot is investigated regardless of HPLC purity.
Charge envelope distortion
If the multi-charge distribution is not Gaussian — gaps in the expected charge states, asymmetric tail, multiple envelopes overlapping — the sample contains more than one species and identity cannot be reported.
Calibration drift > 5 ppm
Calibration is verified against a sodium formate ladder before every batch. If mass accuracy across the scan range exceeds 5 ppm RMS, no client sample is acquired until the instrument is recalibrated.