Home Catalog Quality & COA About Contact Cart
Browse Catalog
Free shipping over $59 · $7.99 US (48h) · $14.99 worldwide (5 days) · Discreet Packaging · Crypto Accepted
HPLC Workflow

Reverse-phase HPLC, at 220 nm, against a reference standard.

Every released lot is profiled on a C18 column with gradient elution, UV detection at 220 nm and quantified as area-percent purity. The chromatogram travels with every shipment.

≥ 99%Release purity floor
220 nmDetection wavelength
C18Reverse-phase column
End-to-end workflow

Six steps from synthesis batch to released lot.

Every peptide goes through the same path. There is no expedited lane, no in-house shortcut, no "this looked fine, ship it." A failure at any step rejects the batch.

  1. 01

    Sample receipt & chain-of-custody

    Lyophilized peptide arrives sealed from the synthesis facility. Lot ID, batch, theoretical sequence and MW are logged. The vial is split into two aliquots: one for analytical, one for archive (retained 24 months).

    ≈ 0.5 h · receiving log
  2. 02

    Reconstitution & dilution

    2 mg of peptide reconstituted in 1 mL of mobile phase A (0.1% TFA in water) with light vortexing. Diluted to 1 mg/mL working concentration. Centrifuged 5 min at 13 000 g to remove particulates that would foul the column.

    ≈ 0.3 h · 1 mg/mL prep
  3. 03

    Column equilibration

    250 × 4.6 mm C18 column (5 µm, 100 Å) equilibrated at 95% A / 5% B, 1.0 mL/min, 30 °C until baseline drift is < 0.0005 AU/min. Pressure stable within ± 5 bar across three blank injections.

    ≈ 0.5 h · until baseline
  4. 04

    Injection & gradient run

    20 µL injection loop, full loop fill. Linear gradient from 5% to 95% acetonitrile (0.1% TFA) over 25 min, 5 min hold at 95%, return to initial in 2 min, re-equilibrate 8 min. Total cycle 40 min.

    ≈ 0.7 h · 40 min cycle
  5. 05

    Detection & integration

    Diode-array UV detector, primary trace at 220 nm (peptide bond), secondary at 280 nm (Trp/Tyr). Peaks integrated by valley-to-valley, baseline corrected, area-percent reported relative to the main product peak.

    automated · integrator
  6. 06

    Release decision

    Main peak ≥ 99.0% area: the lot moves to ESI-MS confirmation. Below 99.0%: the lot is rejected, the synthesis batch is investigated, and the QC officer signs the rejection notice. No exceptions, no rounding.

    QC officer · signed
Anatomy of a chromatogram

Reading what the trace actually says.

A clean release chromatogram is a flat baseline, a sharp main peak, and very little else. Here is how to read every part of one — including the parts most vendor "COAs" hide by cropping.

HPLC · IGF-1 LR3 · Lot LR3-2025-A47 · C18 / 220 nm Main peak · 99.42%
0510152025 min

Main product peak

Sharp, single, high. Area = <b>99.42%</b> of total integrated signal. Retention 13.4 min — matches the reference standard within ± 0.05 min.

Process-related impurities

Three minor peaks visible at 6.6, 13.0 and 16.5 min. Combined area = <b>0.58%</b>. Likely deletion sequences and oxidation byproducts; identified by ESI-MS in the next assay.

Baseline

Flat, low noise (drift < 0.0005 AU/min), no rolling hump from solvent contamination, no late-eluting "garbage" peak after the main product. The chromatogram returns to baseline cleanly.

Method parameters

The exact specs the lab runs against.

If you can't reproduce the assay from the parameters, the COA is opaque. Here is everything you need to repeat the run on your own instrument.

Stationary phase

ColumnC18 reverse-phase, fully end-capped
Dimensions250 mm × 4.6 mm i.d.
Particle size5 µm
Pore size100 Å
Carbon load≈ 17%
Temperature30 °C (column oven)

Mobile phase & gradient

Solvent AH₂O + 0.1% TFA
Solvent BAcetonitrile + 0.1% TFA
Flow rate1.0 mL/min
Gradient5% → 95% B over 25 min
Hold95% B for 5 min
Re-equilibration5% B for 8 min

Detection & integration

DetectorDiode-array UV (DAD)
Primary λ220 nm (peptide bond)
Secondary λ280 nm (Trp / Tyr)
Sampling rate10 Hz
IntegrationValley-to-valley, baseline corrected
ReportingArea-% relative to total

Sample & injection

DiluentMobile phase A
Concentration1 mg/mL working solution
Injection volume20 µL (full-loop)
Replicates3 per lot, ± relative SD reported
System suitabilityUSP <621> criteria, every batch
Reference standardIndependent, lot-traceable
Why 99% is not "basically the same as 95%"

The impurity load doubles between every grade.

A "95% pure" peptide carries ten times the impurity mass of a "99.5% pure" lot. In a 5 mg vial, that's 250 µg of unidentified material vs. 25 µg. In a cell-culture experiment, that difference is the experiment.

95%"research grade"
Impurity load50 mg/g
In a 5 mg vial250 µg

Common floor for grey-market peptides. Identity is plausible; reproducibility is not.

98%generic vendor floor
Impurity load20 mg/g
In a 5 mg vial100 µg

The default release threshold for most internet vendors. Better — but truncations still pass undetected.

99%IGF1 Shop release floor
Impurity load10 mg/g
In a 5 mg vial50 µg

Hard floor. Anything below is rejected and the synthesis batch is investigated. No rounding.

99.5%+typical IGF1 Shop release
Impurity load5 mg/g
In a 5 mg vial25 µg

Where most of our released lots actually land. The 99% line is the floor, not the mode.

Reading the small peaks

What the impurity peaks usually mean.

The minor peaks on a chromatogram are not random noise — each one carries information about how the synthesis went. Here are the patterns we look for and what they signal back to the synthesis facility.

Peak patternLikely causeRetention shiftAction
Single shoulder, leading edge of main peak Diastereomer or D-amino-acid contamination ± 0.1–0.3 min Re-couple if > 0.3%
Sharp peak earlier than main, integer mass shift Truncation (missing residue) − 0.5–2.0 min Reject if > 0.5%
Sharp peak later than main, +16 Da MS shift Methionine / Trp oxidation + 0.3–0.8 min Acceptable if < 0.5%
Tail on main peak, +1 Da MS shift Asn / Gln deamidation + 0.05–0.2 min Investigate if > 0.3%
Late-eluting cluster after the main peak TFA adducts, residual scavengers + 1.0–3.0 min Cosmetic, ≤ 0.2%
Rolling baseline hump under all peaks Solvent contamination, dirty column n/a Re-equilibrate or replace
Hard fails

Four conditions that reject the lot, no override.

The release rule is not a guideline. These are the conditions under which the QC officer signs a rejection notice and the lot returns to the synthesis facility for investigation.

F1

Main peak below 99.0% area

The release floor. Even 98.97% is rejected. There is no "almost passed", no rounding to two significant figures, no "we'll release it as the lower-grade SKU" — the batch is sent back.

F2

Single impurity peak above 0.5%

Even when the main peak clears 99.0%, a single related impurity above 0.5% is treated as a synthesis problem worth investigating. Common cause: a truncation that the gradient didn't fully separate.

F3

Retention time off > 0.2 min vs. reference

If the main peak elutes at the wrong time relative to the lot-traceable reference standard, the molecule has likely changed — either a sequence error or a major modification. Sample is held pending MS investigation.

F4

System suitability failure

USP <621> criteria — theoretical plates, tailing factor, capacity factor, signal-to-noise — are checked on a system suitability standard before every batch. If the column or detector is drifting, no sample is run until it's restored.

FAQ

What researchers ask about HPLC purity.

Why detect at 220 nm instead of 280 nm?
220 nm is the absorption band of the peptide bond itself, so every residue contributes signal proportionally to its content. 280 nm only sees Trp, Tyr and (weakly) Cys — fine for protein quantification, but blind to peptides that lack aromatic residues. We log both wavelengths and report 220 nm as the canonical purity number.
Could two impurities co-elute under the main peak?
In principle, yes — and this is exactly why HPLC is paired with ESI-MS. A co-eluting truncation or modification would inflate the area-percent number but produce a secondary ion in the mass spectrum. Our release requires both assays to clear, not just one.
Why use TFA in the mobile phase?
TFA at 0.1% acts as an ion-pairing modifier — it neutralises the basic side-chains (Lys, Arg, His) and produces sharp, symmetric peaks instead of broad tails. The downside is some MS suppression, which is why the same sample is re-run for ESI-MS in a TFA-compatible source rather than relying on the HPLC-MS hyphenation.
What's a "system suitability" check?
A short pre-batch run (typically 5–6 injections of a reference standard) used to confirm the instrument is performing within USP <621> specification: theoretical plates > 2000, peak tailing < 1.5, capacity factor 2–10, signal-to-noise > 50. If any criterion fails, no client sample is injected until the column or detector is restored.
Are the chromatograms in your COAs real or stylised?
Real. Every COA we publish embeds the actual integrator output for that specific lot — peak retention times, areas, percentages and notes. The lab signs the file and we ship a copy unmodified with the order. The "anatomy" diagram on this page is a stylised SVG; the chromatograms in the COAs are not.
Do you publish the lab's accreditation certificate?
Yes — institutional buyers can request the ISO/IEC 17025:2017 scope certificate, the method validation reports, and the proficiency testing results from the most recent round. Email us and we'll forward them under NDA.